Brandtech-scientific Transferpette-8/-12 electronic Manual do Utilizador Página 59

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 58
English
Final test values related to the nominal capacity (maximum volume)
and the indicated volume steps indicated on the instrument, obtained
when instrument and distilled water are equilibrated at ambient tem-
perature (20 °C/68 °F) and with smooth operation. According to
DIN EN ISO 8655.
Volume range Volume step A* CV* Increment Recommended
µl µl ± % % µl type of tip, µl
0.5 - 10 10 1.2 0.8 0.01 0.5 - 20
5 2.0 1.5
1 8.0 4.0
1 - 20 20 1.0 0.5 0.02 0.5 - 20
10 2.0 1.0
2 8.0 3.0
5 - 100 100 0.8 0.25 0.1 2 - 200
50 1.6 0.4
10 4.0 1.5
10 - 200 200 0.8 0.25 0.2 2 - 200
100 1.4 0.4
20 4.0 1.3
15 - 300 300 0.6 0.25 0.5 5 - 300
150 1.2 0.4
30 3.0 1.2
Accuracy tolerances for the Transferpette
®
-8/-12 electronic
* A
= Accuracy, CV
= Coefficient of Variation
Accuracy Table
!
20 °C
Ex
59
Testing instructions (SOPs) and a demo version of the EASYCAL
4.0 calibration software are available
for download at www.brand.de.