Accuracy Table
Technical Data
PD-Tips from BRAND, 20 °C 'Ex',
PD-Tip
size
Volume
range (µl)
A*
<
± %
Stroke setting
=
%
of nominal volume
CV*
<
%
Stroke setting
=
%
of nominal volume
1
=
2% 3
=
6% 5
=
10% 1
=
2% 3
=
6% 5
=
10%
0.1 ml 2 - 10 4.0 2.4 1.6 6.0 3.0 2.0
0.5 ml 10 - 50 2.5 1.5 1.0 2.5 1.5 1.0
1.0 ml 20 - 100 2.5 1.5 1.0 2.0 1.2 0.8
1.25 ml 25 - 125 2.5 1.4 0.9 2.0 1.1 0.7
2.5 ml 50 - 250 1.8 1.1 0.7 1.5 0.9 0.6
5.0 ml 100 - 500 1.8 1.1 0.7 1.5 0.9 0.7
10.0 ml 200 - 1000 1.8 1.1 0.7 2.0 1.2 0.8
12.5 ml 250 - 1250 1.8 1.1 0.8 3.2 2.0 1.4
25.0 ml 500 - 2500 1.5 0.9 0.6 3.0 1.5 1.0
50.0 ml 1000 - 5000 1.5 0.8 0.5 5.0 1.8 1.2
A* = Accuracy, CV* = Coefficient of variation
The HandyStep
®
S
can also be operated with compatible tips from other
manufacturers. The user must check their compatibility and perform a
function and volume test before use!
Important!
35
Error limits refer to the dispensing volume set relative to the PD-Tip, obtained when instrument,
PD-Tip and distilled water are equilibrated at ambient temperature (20 °C/68 °F) and with
smooth operation. The testing is according to DIN EN ISO 8655-5.
The nominal volume is the maximum volume printed on the PD-Tip.