Brandtech-scientific macro Manual do Utilizador Página 27

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 26
8
6
English
27
8. The entire instrument incl. membrane
filter is autoclavable at 121 °C (2 bar)
according to DIN EN 285.
For autoclaving remove suction bellow.
6. Place the new membrane filter with
the thicker end facing downwards into the
adapter (with the smaller conical side up).
Not every membrane filter is suitable. Only use genuine manufacturer’s
recommended accessories. Filters of a pore size of 3 µm are a standard
equipment of the instrument. The membrane filter is autoclavable up
to 5 times at 121 °C. It is the user's responsibility to ensure effective
autoclaving.
Note:
Cleaning · Autoclaving
7. Screw in the adapter support hand tight.
Vista de página 26
1 2 ... 22 23 24 25 26 27 28 29 30 31 32 ... 83 84

Comentários a estes Manuais