Brandtech-scientific Transferpette S -8/-12 Manual do Utilizador Página 51

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 50
English
51
Troubleshooting
Problem Possible cause Corrective action
Tip dripping
(instrument leaks)
Unsuitable tip Only use high-quality
tips
Tip not seated tightly Press tip on firmly
The instrument does
not aspirate or as-
pirates too little; the
discharged volume
is too low.
Seal contaminated Clean seal
The seal or cone is
damaged
Replace seal or shaft
The piston is contami-
nated or damaged
Clean or replace piston
Aspiration is
too slow
Shaft clogged Clean shaft
Discharged volume
is too large
Pipetting button
pressed too far into
the blow-out position
before sample uptake
Operate properly.
See 'Pipetting',
page 34.
Piston is difficult
to move
The piston is contami-
nated or needs oil
Clean and oil the piston
Vista de página 50
1 2 ... 46 47 48 49 50 51 52 53 54 55 56 ... 132 133

Comentários a estes Manuais

Sem comentários