Brandtech-scientific PLT unit Pipette Leak Testing Unit Manual do Utilizador Página 46

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 45
1. Unscrewing the adapter lock nut.
2. Remove the pipette adapter.
3. If the PE filter is clogged with moisture or contaminants, it must be
changed. In this case, carefully withdraw the PE filter from the bottom
end of the silicone adapter.
4. Insert a new filter.
5. Insert the selected pipette adapter into the adapter housing (in the figure,
Adapter 2 for pipettes without tips).
6. Screw the lock nut down firmly.
7. After changing the adapter, perform the Quick Test.
Test preparation
Changing the adapter and filter
1
2
3
4
5
6
The 4-channel adapter consists of an adapter with
an integrated lock nut. The entire ensemble is
exchanged.
Note:
46
Vista de página 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 131 132

Comentários a estes Manuais

Sem comentários