Error Limits
Error limits related to the nominal capacity (= maximum volume) indicated on
the instrument, obtained when instrument and distilled water are equilibrated
at ambient temperature (20 °C/68 °F). Testing takes place according DIN EN
ISO 8655-6 with a completely filled instrument and with uniform and smooth
dispensing.
Note:
* A = Accuracy, CV = Coefficient of Variation
Error limits Dispensette
®
The error limits are well within the limits of DIN EN ISO 8655-5. The maxi-
mum error limit for a single measurement can be calculated EL = A + 2 CV
(e.g. for volume 25 ml: 125 µl + 2 x 25 µl = 175 µl).
Nominal volume
ml
A* ≤ ±
%
µl
CV* ≤
%
µl
0,5 1.0 5 0.2 1
1 0.5 5 0.1 1
2 0.5 10 0.1 2
5 0.5 25 0.1 5
10 0.5 50 0.1 10
25 0.5 125 0.1 25
50 0.5 250 0.1 50
100 0.5 500 0.1 100
Partial volume
The percentage values for A and CV are relative to the nominal volume (V
N
)
and must be converted for partial volumes (V
T
).
V
N
V
T
A
T
= · A
N
e.g. volume A* ≤ ±
% µl
CV* ≤
% µl
V
N
25.0 0.5 125 0.1 25
V
T
= 50% N 12.5 1.0 125 0.2 25
V
T
= 10% N 2.5 5.0 125 1.0 25
* A = Accuracy, CV = Coefficient of Variation